RTD temperature probes
Ex „i“ for use in areas
with an explosion hazard (Ex areas)
Persons concerned:
Experienced professional electricians
as per EU Directive 1999/92/EC
and trained personnel
B 902820.0.2
B 902821.0.2
B 902xxx.0.2
Operating Manual
2013-01-15/00404304
Page view 0
1 2 3 4 5 6 ... 11 12

Summary of Contents

Page 1 - RTD temperature probes

RTD temperature probesEx „i“ for use in areaswith an explosion hazard (Ex areas)Persons concerned:Experienced professional electriciansas per EU Direc

Page 2

8 Connection circuits for RTD temperature probes (applies to both head-mounted and cable-connected RTDs from JUMO) 2-wire circuit 3-wire circuit

Page 4

JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc.Street address:Moritz-Juchheim-Straße 136039 Fulda, GermanyDelivery address:

Page 5 - Linear (surface temperature)

Contents 1 Equipment and intended application... 2 2 Marki

Page 6

- 2 - 1 Equipment and intended application RTD temperature probes from JUMO are used as intrinsically safe and/or flameproof equipment for tempera

Page 7

- 3 - 4 Conformance with standards WARNING: This operating manual is only valid for the specified test certificate and the norms that are associat

Page 8

Temperature class Max. surface temperature for the equipment1 Ignition temperature of the flammable materials T1 450 °C  450 °C T2 300 °C  300  45

Page 9

- 5 - Example 1 A thermometer is to be used in temperature class T4 (maximum temperature 135 °C, safety reduction of the limit: 5 °C). Thermowell

Page 10

- 6 - It is not mandatory to connect the versions in flameproof enclosure to intrinsically safe circuits, however, it must always be ensured that th

Page 11

The ignition or smoldering temperature of the dust or dust-air mixture that is present must be defined or determined by the operator of the plant or s

Page 12

- 8 - 6 Installation The valid European and national regulations must be observed for installation and operation. The generally accepted state of